Chemetron Medical Gas Outlets with Box
Medical gas outlets are used for medical gas system daily medical care and installed at gas systems delivery points(wardroom bed head units, medical pendants, and equipment wall) to provide the required medical gas.
Product Detail
Chemetron medical gas outlets include wall mount type and rough-in assembly type. Wall mount type is using installed on the wall, available in plastic and metal covers. Rough-in assembly type with a plastic decorative cover is using on the bed head unit.
AmcareMed Chemetron gas outlets can provide 7 kinds of gas, including oxygen, medical air, vacuum, nitrous oxide, carbon dioxide, nitrogen, and instrument air. Different gases can be equipped with corresponding gas adapters.
Specifications
- Complies with NFPA 99 and CGA Standard
- 360° swivel inlet pipe for easy installation
- Insertion and extraction 10000 times without failure
- Working pressure: 4-8bar
- O.D. Size: 8mm, 10mm, 12mm K type copper pipeline
- Accepts only Chemetron gas connection fitting
- 100% pressure and leak-tested, can be used safely and securely
- The outlet with two-stage one-way valve for easy maintenance without cutting gas supply
- NFPA full-color standard for easy gas identification and the gas name is marked on the outlet bracket and chrome-plated brass outlet valve body, avoid installation or use errors
Model
Ordering Information | ||
Gas Service | Wall mounting | Bedhead (console) mounting |
OXYGEN | CH-WAL-OXY | CH-CON-OXY |
MEDICAL AIR | CH-WAL-MA | CH-CON-MA |
MEDVAC | CH-WAL-VAC | CH-CON-VAC |
NITROUS OXIDE | CH-WAL-N2O | CH-CON-N2O |
NITROGEN | CH-WAL-NIT | CH-CON-NIT |
CARBON DIOXIDE | CH-WAL-CO2 | CH-CON-CO2 |